Personlized Products Touch Screen Module - AC/DC Medical Adapter ADA360K240S001A – DWIN Detail:
Video
Specification

Selection
| Max. Power | 36W |
| Input voltage | 100-240 VAC |
| Voltage/Current | DC24V /1500mA |
| Efficiency | 90% |
| Max. Capacitive Load | 2000 uF |
| Standby Power Consumption | 0.075W |
Input Parameters
| Input Voltage Range | AC input: 100-240VAC |
| Input Frequency | 50-60Hz |
| Input Current | 115VAC, Max. 0.7A 230VAC, Max. 0.5A |
| Impulse Current | 115VAC, Typic: 10A 230VAC, Typic: 25A |
| Leakag current | 264VAC/60Hz, typic: 0.1mA |
Output Paremeters
| Output Voltage Accuracy | Typic: ±5% |
| Linear Regulation Rate | Full load, typic: ±0.5% |
| Load Regulation Rate | Typic: ±1% |
| Ripple Noise | 20MHz, Typic: 50mV, Max. 100mV |
| Temperature Drift Coefficient | Typic: ± 0.02%/℃ |
| Short Circuit Protection | Long term short circuit, self recovery |
| Over Current Protection | ≥130% IO self recovery |
| Over Voltage Protection | 12V output: ≤27VDC 24V output: ≤40VDC |
| Standby Power | Max. 0.075W |
| Minimum Load | Min. 0% |
| Power Down Hold Time | 115VAC input: typic 15ms 230VAC input: typic 80ms |
General Parameters
| Isolation Voltage(Input/Output) | Min. 4000VAC(The test time is 1 minute, and the leakage current is ≤ 5mA) |
| Orerating Temp. | -40/+70℃ |
| Storage Temp. | -40/+85℃ |
| Storage Humidity | 95%RH |
| Wave Soldering | 260±5℃; time: 5-10S |
| Manual Welding | 360±10℃; time: 3-5S |
| Switching Frequency | 65kHz |
| Safety Standards | EN60601-1 |
| Safety Certification | EN60601-1 |
| Safety Level | 2XMOPP(Primary side-Secondary side) |
| Mean Time Between Failure | ≥300000H (25%) |
EMC Parameters
| Conducted Disturbance | EN55011(CISPR11)/EN55032(CISPR32) CLASSB |
| Radiation Disturbance | EN55011(CISPR11)/EN55032(CISPR32) CLASSB |
| Voltage Flicker | EN61000-3-2 |
| Harmonic Current | EN61000-3-2 |
| ESD | ±8KV/Contact ±2KV, 4KV, 8KV, 15KV/Air |
| Radiated Immunity | 10V/m |
| Burst Immunity | ±2KV |
| Surge Immunity | ±0.5KV/±1KV |
| Conducted Disturbance Immunity | 3Vm/0.15MHz-80MHz 6Vm(Within 15m band) /0.15MHz-80MHz 80%AM,1kHz |
| Immunity to voltage sags, sags and short interruptions | 0%UT 0.5cycle 0°,45°,90°,135°,180°, 270°,315°。 0%UT 1cycle 70%UT 25/30 cycle 0° 0%UT 250/300 cycle |
Application
Product detail pictures:

Related Product Guide:
We have now numerous great personnel members good at advertising, QC, and working with kinds of troublesome dilemma from the creation course of action for Personlized Products Touch Screen Module - AC/DC Medical Adapter ADA360K240S001A – DWIN, The product will supply to all over the world, such as: Uruguay , France , Bolivia , We always hold on the company's principle honest, professional, effective and innovation, and missions of: let all drivers enjoy their driving at night, let our employees can realize their value of life, and to be stronger and service more people. We are determined to become the integrator of our product market and one-stop service provider of our product market.