Personlized Products Touch Screen Module - AC/DC Medical Adapter ADA360K240S001A – DWIN Detail:
Video
Specification
Selection
Max. Power | 36W |
Input voltage | 100-240 VAC |
Voltage/Current | DC24V /1500mA |
Efficiency | 90% |
Max. Capacitive Load | 2000 uF |
Standby Power Consumption | 0.075W |
Input Parameters
Input Voltage Range | AC input: 100-240VAC |
Input Frequency | 50-60Hz |
Input Current | 115VAC, Max. 0.7A 230VAC, Max. 0.5A |
Impulse Current | 115VAC, Typic: 10A 230VAC, Typic: 25A |
Leakag current | 264VAC/60Hz, typic: 0.1mA |
Output Paremeters
Output Voltage Accuracy | Typic: ±5% |
Linear Regulation Rate | Full load, typic: ±0.5% |
Load Regulation Rate | Typic: ±1% |
Ripple Noise | 20MHz, Typic: 50mV, Max. 100mV |
Temperature Drift Coefficient | Typic: ± 0.02%/℃ |
Short Circuit Protection | Long term short circuit, self recovery |
Over Current Protection | ≥130% IO self recovery |
Over Voltage Protection | 12V output: ≤27VDC 24V output: ≤40VDC |
Standby Power | Max. 0.075W |
Minimum Load | Min. 0% |
Power Down Hold Time | 115VAC input: typic 15ms 230VAC input: typic 80ms |
General Parameters
Isolation Voltage(Input/Output) | Min. 4000VAC(The test time is 1 minute, and the leakage current is ≤ 5mA) |
Orerating Temp. | -40/+70℃ |
Storage Temp. | -40/+85℃ |
Storage Humidity | 95%RH |
Wave Soldering | 260±5℃; time: 5-10S |
Manual Welding | 360±10℃; time: 3-5S |
Switching Frequency | 65kHz |
Safety Standards | EN60601-1 |
Safety Certification | EN60601-1 |
Safety Level | 2XMOPP(Primary side-Secondary side) |
Mean Time Between Failure | ≥300000H (25%) |
EMC Parameters
Conducted Disturbance | EN55011(CISPR11)/EN55032(CISPR32) CLASSB |
Radiation Disturbance | EN55011(CISPR11)/EN55032(CISPR32) CLASSB |
Voltage Flicker | EN61000-3-2 |
Harmonic Current | EN61000-3-2 |
ESD | ±8KV/Contact ±2KV, 4KV, 8KV, 15KV/Air |
Radiated Immunity | 10V/m |
Burst Immunity | ±2KV |
Surge Immunity | ±0.5KV/±1KV |
Conducted Disturbance Immunity | 3Vm/0.15MHz-80MHz 6Vm(Within 15m band) /0.15MHz-80MHz 80%AM,1kHz |
Immunity to voltage sags, sags and short interruptions | 0%UT 0.5cycle 0°,45°,90°,135°,180°, 270°,315°。 0%UT 1cycle 70%UT 25/30 cycle 0° 0%UT 250/300 cycle |
Application
Product detail pictures:

Related Product Guide:
We keep on with the basic principle of quality to start with, support very first, continuous improvement and innovation to meet the customers for your management and zero defect, zero complaints as the quality objective. To great our service, we offer the items with all the superior top quality at the reasonable selling price for Personlized Products Touch Screen Module - AC/DC Medical Adapter ADA360K240S001A – DWIN, The product will supply to all over the world, such as: Mecca , Lesotho , Iran , Company name, is always regarding quality as company' s foundation, seeking for development via high degree of credibility , abiding by ISO quality management standard strictly, creating top-ranking company by spirit of progress-marking honesty and optimism.