AC/DC Medical grade Wide input voltage High reliability Small size Low no-load voltage Power Module: ADM450K150S001A

Features:

● Wide input voltage: the input working voltage range is 100-240VAC.
● Low power consumption: No-load <0.075W.
● High Energy Efficiency: Six levels of energy consumption, power efficiency up to 88%.
● High reliability: Comply with EN60601-1 CLASS Ⅱ safety level and 2×MOPP insulation protection level, and pass CE.
● Flame retardant insulation: UL94V-0 flame retardant heat resistant material.
● Protection types: short circuit protection, over current protection, over voltage protection, and self-recovery.
● Small size: 115.25*52.20*31.50mm, it can be put into products with high volume requirements.

 


Specification

Product Tags

Product detail
03
Input Parameter
Voltage Range 100~240 VAC,100-370VDC
Frequency Range 47~63Hz
Input Current (Max.) 1.1A @ 100VAC,0.9A @ 240VAC
Efficiency (Typ.) 88%
Standby Consumption 0.1W
Impulse Current (Typ.) 30A @ 100VAC,40A @ 240VAC
Leakage Current (Typ.) 0.1mA @ 264VAC, 60Hz
Output Parameter
Output Voltage 15VDC
Voltage Tolerance ±5%
Output Current 3000mA
Rated Power (Max.) 45W
Line Regulation ±3% at full load
Max. Capacitive Load 7000uF
Load Regulation ±3%
Ripple & Noise 90mV (Typ.), 150mV (Max.)@20MHz
Frequency (Typ.) 65kHz
Hold up Time (Typ.) 15mS @ 100VAC

80mS @ 240VAC

Environment
Operating Temperature -40 ~ +70℃
Storage Temperature -40 ~ +85℃
Storage Humidity 95%RH (Max.)
Power Derating 3.6%/℃ @ -40 ~ -25℃

2.4%/℃ @ +50 ~ +70℃,15VDC

2.7%/℃ @ +55 ~ +70℃

Temperature Coefficient ±0.02%/℃
Soldering Temperature 260±5℃ @ Wave Soldering, 5~10s

360±10℃ @ Manual Soldering, 3-5s

Protection Function
Short Circuit Long term short circuit, auto recovery.
Over Current ≥140%IO auto recovery
Over Voltage ≤23VDC @ 15V output
IEC Safety Class CLASSII
Electric Shock Protection 2×MOPP @ primary to secondary
Reliability
MTBF ≥300000H @ 25℃
Safety Standards / Directives
Medical safety EN60601-1
CE Compliant
Isolation Voltage (Min.) Input to Output 4000VAC @ 1 minute test and the leakage current is smaller than 5mA.
EMC
 

 

EMI

 

 

 

 

Parameter Standard Test Level / Note
Conducted emission EN55011(CISPR11)/EN55032(CISPR32) CLASS B
Radiated emission EN55011(CISPR11)/EN55032(CISPR32) CLASS B
Voltage flicker EN61000-3-2 -
Harmonic current EN61000-3-2 -
EMS

 

 

 

 

 

 

Parameter Standard Test Level / Note
Electrostatic Discharge IEC/EN61000-4-2 ±8KV/Contact

±2、4、8、15KV/Air

Radiate Susceptibility IEC/EN61000-4-3 10V/m
Electrical Fast Transient burst IEC/EN61000-4-4 ±2 KV
Surge IEC/EN61000-4-5 ±0.5/±1 KV

Professional/Family medicine

Conducted Susceptibility IEC/EN61000-4-6 3Vm/0.15MHz-80MHz 6Vm

(Within15m band)

0.15MHz-80MHz

80%AM,1kHz

Voltage Dips and Interruption IEC/EN61000-4-11 0%UT 0.5cycle

0°,45°,90°,135°,180°,270°,315°。

0%UT 1cycle

70%UT 25/30 cycle 0°

0%UT 250/300 cycle

 

Mechanical Specification
Dimension 115.25*52.2*31.5mm
Enclosure Material Black flame retardant and heat resistant plastics(UL94V-0)
Cooling mode Natural air cooling