Specification
 
		    
			 
		    
			| Voltage Range | 100~240 VAC | 
| Frequency Range | 50~60Hz | 
| Input Current (Max.) | 0.7A @ 115VAC,0.5A @ 230VAC | 
| Efficiency (Typ.) | 90% | 
| Standby Consumption | 0.075W | 
| Impulse Current (Typ.) | 10A @ 115VAC,25A @ 230VAC | 
| Leakage Current (Typ.) | 0.1mA @ 264VAC, 60Hz | 
| Output Voltage | 24VDC | 
| Voltage Tolerance | ±5% | 
| Output Current | 3000mA | 
| Rated Power (Max.) | 36W | 
| Line Regulation | ±0.5% at full load | 
| Max. Capacitive Load | 5000uF | 
| Load Regulation | ±1% | 
| Ripple & Noise | 50mV (Typ.), 100mV (Max.)@20MHz | 
| Frequency (Typ.) | 65kHz | 
| Hold up Time (Typ.) | 15mS @ 115VAC80mS @ 230VAC | 
| Operating Temperature | -40 ~ +70℃ | 
| Storage Temperature | -40 ~ +85℃ | 
| Storage Humidity | 95%RH (Max.) | 
| Power Derating | 2.7%/℃ @ -40 ~ -25℃ 2.4%/℃ @ +50 ~ +70℃,24VDC 2.7%/℃ @ +55 ~ +70℃ | 
| Temperature Coefficient | ±0.02%/℃ | 
| Soldering Temperature | 260±5℃ @ Wave Soldering, 5~10s 360±10℃ @ Manual Soldering, 3-5s | 
| Short Circuit | Long term short circuit, auto recovery. | 
| Over Current | ≥130%IO auto recovery | 
| Over Voltage | ≤40VDC @ 24V output | 
| IEC Safety Class | CLASSII | 
| Electric Shock Protection | 2×MOPP @ primary to secondary | 
| MTBF | ≥100000H @ 25℃ | 
| Medical safety | EN60601-1 | |
| CE | Compliant | |
| Isolation Voltage (Min.) | Input to Output | 4000VAC @ 1 minute test and the leakage current is smaller than 5mA. | 
| EMI | Parameter | Standard | Test Level / Note | 
| Conducted emission | EN55011(CISPR11)/EN55032(CISPR32) | CLASS B | |
| Radiated emission | EN55011(CISPR11)/EN55032(CISPR32) | CLASS B | |
| Voltage flicker | EN61000-3-2 | - | |
| Harmonic current | EN61000-3-2 | - | |
| EMS | Parameter | Standard | Test Level / Note | 
| Electrostatic Discharge | IEC/EN61000-4-2 | ±8KV/Contact±2、4、8、15KV/Air | |
| Radiate Susceptibility | IEC/EN61000-4-3 | 10V/m | |
| Electrical Fast Transient burst | IEC/EN61000-4-4 | ±2 KV | |
| Surge | IEC/EN61000-4-5 | ±0.5KV/±1 KVProfessional/Family medicine | |
| Conducted Susceptibility | IEC/EN61000-4-6 | 3Vm/0.15MHz-80MHz6Vm(Within 15m band)/0.15MHz-80MHz 80%AM,1kHz | |
| Voltage Dips and Interruption | IEC/EN61000-4-11 | 0%UT 0.5cycle 0°,45°,90°,135°,180°,270°,315°。 0%UT 1cycle 70%UT 25/30 cycle 0° 0%UT 250/300 cycle | 
| Dimension | 70*47*48mm | 
| Enclosure Material | Black flame retardant and heat resistant plastics(UL94V-0) | 
| Cooling mode | Natural air cooling | 
| Power cable specifications | ul2464 × 22Awg * 1m, 5.5 * 2.5 straight head + magnetic ring + SR | 
DMG48480C021_03W (Commercial Grade)



 
            